Back to search results

EQUILIBRIUM STEAM DISPENSER DRP-10(DRP-13)

The equilibrium steam dispenser "DRP-10 (DRP-13)" is designed for the introduction of volatile components from matrices into the chromatograph, the introduction of which into the chromatograph is impossible or undesirable. Examples of such matrices are natural, drinking and waste water, soil, biological fluids, food and beverages, various polymers, building materials, pharmaceutical products, etc. The use of DRP-10 (DRP-13) makes it possible to avoid contamination of the evaporator, column and chromatograph detector with non-volatile or non-volatile substances, to avoid the consumption of solvent required for extraction, to increase the reproducibility of analyses.
Technical specifications
The temperature of the thermostat of containers with a breakdown is from ambient temperature to 150 °C.
The error of maintaining the temperature is not more than 0.1 ° C.
The volume of the container with the sample is 20 ml, it is possible to use vials from 6 to 40 ml.
The number of simultaneously thermostatically controlled containers is 4.
The dosed volume of the equilibrium vapor is up to 2 ml.
The temperature of the syringe is from ambient temperature to 150 °C.
The gas consumption for purging is from 5 to 300 ml/min.
Electric power supply from the AC network with voltage (220index) V, frequency (50 ± 1) Hz.
Overall dimensions (width × depth × height) — 350×270×170 mm.
Power consumption — no more than 360 VA.
Weight — no more than 6.5 kg.