Back to search results

LOGARITHMIC DIGITAL CURRENT METER (ITLC)

Back to search results
LOGARITHMIC DIGITAL CURRENT METER (ITLC)
Manufacturer:
NPF "META-KHROM"
Location:
Yoshkar-Ola
Applying:
The hardware and software complex for measuring the signals of chromatographs "Meta-chrome" (complex) is designed to automate the old fleet of chromatographs of the type LHM, model-3700, series "Color", etc. in terms of measuring the chromatographic signal and automating the processing of chromatographic information. The complex consists of two interconnected functional modules: a logarithmic digital current meter (ITLC) and a chromatographic information processing program "NetChrom" (hereinafter referred to as the processing program).
technical specifications
The number of analog inputs is 1.
Data exchange with a personal computer — via the standard RS-232C interface.
The input impedance is at least 1×106 Mohm.
The range of input currents is from 1×10-14 to 5×10-7 A.
The minimum measured half—width of the chromatographic peak is 0.5 s.
The level of fluctuation noise brought to the input is no more than 1.2 × 10-14 A.
The reduced error of the linearity of the signal conversion, no more than ± 1%.
Electric power supply from a single—phase AC network with voltage from 187 to 242 V, frequency (50 ± 1) Hz, operating mode - continuous.
Power consumption (without a personal computer), no more than — 3 VA.
Overall dimensions (width × depth × height), not more than — 230 ×150 × 55 mm.
Weight, not more than 1.3 kg.
Reviews