Back to search results

MARK-409T Dissolved Oxygen Analyzer

Back to search results
MARK-409T Dissolved Oxygen Analyzer
Manufacturer:
VZOR
Location:
Nizhny Novgorod
Measurement of dissolved oxygen concentration and temperature of aqueous media,
including deaerated ones.Convenience and accuracy of measurement |
The measurement error is ±(0.001 + 0.035 *A) mg/dm3, (A is the measured value).
The ability to work on small flows (from 25 ml / min). Routine maintenance 1 time per year.
Highly stable sensor |
High reaction speed. Sealed heavy-duty hydrophobic membrane.
Increased mechanical strength of the structure.
Two channels | Freely programmable measurement ranges.
The possibility of independent measurements at two points.
The possibility of placing the converter unit at a remote distance
from the sampling point | Up to 100 meters.
Communication with external devices | Galvanically isolated current outputs 0-5/4-20/0-20 Ma.
RS 485 port. MODBUS RTU exchange protocol.
Hydropanel GP 409 T |
The possibility of placing the measuring complex on a single panel.
The supply lines are stainless steel.
Routine maintenance and calibration of the sensor – without interrupting the flow.
Continuous monitoring of water-chemical regimes at thermal and nuclear power facilities.
Reviews
There are no entries.