Back to search results

PYROLYZER

The pyrolytic gas chromatography method is widely used in the study of polymer materials, plastics, rubbers, rubbers, paints and other high-molecular compounds of natural and artificial origin. The method is based on thermal decomposition of sample samples in an inert medium (without oxygen access), followed by chromatographic separation and analysis of the formed volatile compounds. Thus, the resulting pyrogram allows us to judge the composition of the original sample. Pyrolysis is also used in geochemical studies.
Technical specifications
The volume of the capsule is 56 ml.
Pyrolysis temperature — from 200 to 1000 °C.
The error of maintaining the temperature is not more than 1.0 ° C.
The heating time of the sample to the pyrolysis temperature is no more than 2 seconds.
The cooling time of the capsule when moving from the hot zone to the cold zone after pyrolysis is no more than 10 seconds.
The time of continuous operation in the pyrolysis mode is no more than 60 seconds.
Electric power supply from the AC network with voltage (220index) V, frequency (50 ± 1) Hz.
Power consumption — no more than 100 VA.
Overall dimensions of the control unit (width ×depth×height) — 210×190×100 mm.
Overall dimensions of the input node (width× depth×height) — 90×135×160 mm.
The weight of the control unit is not more than 2.5 kg.
The weight of the input node is not more than 0.9 kg.
Reviews