Air Purification Units BOV-1
They are designed for purification of nitrogen (BOV-1A) and air (BOV-1) from hydrocarbons
Technical specifications
Inlet air pressure is not more than 825 ± 5% kPa (8 ati)
Output pressure control limits is from 0 to 413 kPa (from 0 to 4.0 ati)
Minimum required excess of inlet pressure over outlet pressure is not less than 103 kPa(1 ati)
Maximum permissible air flow is not more than 1.0 l/min
Concentration of hydrocarbons at maximum flow rate in the outlet air is no more than 0.1 ppm
Water vapor concentration at 20°C and 100 kPa is no more than 10 ppm
Instability of the output pressure when the flow rate changes from 0 to the maximum value is not worse than ± 7 kPa
Instability of the output pressure when the input pressure changes by ± 100 kPa is not worse than ± 7
Ready time after switching on at a flow rate not exceeding the maximum value is no more than 30 min
Power consumption is not more than 50 VA
Overall dimensions, (width x depth x height) are no more than 185x310x310 mm
Weight is not more than 5.0 kg
Working conditions:
ambient temperature, °C, from +10 to +35
relative humidity of the ambient air, %, no more, at 35 ° C 80
electric power from a single-phase AC network
with a voltage of 220 +10% -15%
and a frequency of 50 ± 1 Hz
Produced in: Moscow